|
|
|
|
|
Session 125
Poster Abstracts Viral Load Assays Thursday, 1:30 - 3:30 pm Hall A |
Background: Viral load testing is the standard of care in
monitoring response to hepatitis C virus (HCV) therapy in HCV/HIV co-infected
individuals. Real-time PCR assays hold great promise for this application
because they are very sensitive and have a broad linear range. However, the
comparative performance of 2 commercial real-time viral load assays, the Abbott
TaqMan ASR (Abbott) and the Roche TaqMan
RUO (Roche) has not been evaluated. We report a multi-laboratory comparison on
an evaluation of these two assays, to determine the limit of detection, linear
range, reproducibility, and agreement.
Methods: Plasma was obtained from an HCV-infected
individual (genotype 1b) and the viral load was determined using the Roche Cobas Amplicor HCV Monitor test. This
stock material was diluted in normal plasma to concentrations of 1 to 7 log10
IU/mL. Each of 4 panels had 7 replicates of each concentration,
which was tested in 3 (Abbott) or 4 (Roche) labs. Data were log10
transformed prior to analysis.
Results: Both assays detected all 28 replicates with a
concentration of 1 log10 IU/mL and were
linear to 7 log10 IU/mL. Overall the Roche
assay was more reproducible than the Abbott assay, and both assays were less
reproducible at lower concentrations of virus. For the Roche assay, the standard
deviation (SD) was 0.18 and 0.09 log10 for replicates of 2 and 5.5
log10 IU/mL, respectively. While the SD
for the Abbott assay was 0.42 and 0.26 log10 for replicates of 2 and
5.5 log10 IU/mL, respectively. For the
Roche assay, the within-lab SD was below 0.15 log10 for all 4 labs
at viral load viral values of ≥ 2.5 log10 IU/mL. For the Abbott assay, the within-lab SD ranged from
0.17 to 0.43 log10 at viral load values > 2.5 log10
IU/mL; for 3 of the 4 labs the SD was ≤ 0.15
log10 for viral load values ≥ 5.5 log10 IU/mL. In general, the viral load values obtained with the
Roche assay were 0.2 to0.7 log10 greater than those obtained with
the Abbott assay.
Conclusions:
The Roche assay can reliably detect
5-fold changes in viral load for values ≥ 2.5 log10 IU/mL, while this level of reliability is reached with the
Abbott assay at values > 5.5 log10 IU/mL.
Overall the sensitivity and linear range (at least 6 log10) of these
real-time assays enables them to be used for both HCV diagnostics as well as therapeutic
monitoring, which is not possible with the current RT-PCR assays.
Keywords: HCV; viral load; real-time assays
![]() |